MS30 Series multi angles spectrophotometer for car painting

| Model | MS3012 | MS3008 | MS3006 | MS3005 | MS3003 | MS3000 |
| Measurement Geometry | 12 measurement angles (7 illumination sources, 2 receivers) | 8 measurement angles (6 illumination sources, 2 receivers) | 6 measurement angles (6 illumination sources, 1 receivers) | 5 measurement angles (5 illumination sources, 1 receivers) | 3 measurement angles (3 illumination sources, 1 receivers) | 3 measurement angles (3 illumination sources, 1 receivers) |
| Measure Angle | 45° Receiver: | |||||
| 45as-15°,45as15°,45as25°,45as45°,45as75°,45as110° | 45as-15°,45as15°,45as25°,45as45°,45as75°,45as110° | 45as-15°,45as15°,45as25°,45as45°,45as75°,45as110° | 45as15°,45as25°,45as45°,45as75°,45as110° | 45as25°,45as45°,45as110° | 45as15°,45as45°,45as110° | |
| 15°Receiver: | 15°Receiver: | / | / | / | / | |
| 15as-45°,15as-15°,15as15°,15as-30°,15as45°,15as80° | 15as-45°,15as-15° | / | / | / | / | |
| Conform to the standards:ASTM D 2244,E 308,E 1164,E 2194, E2539,DIN 5033,5036,6174,6175-1,6175-2;ISO 7724, 11664-4 SAE J 1545 | ||||||
| Application | Provide accurate and consistent color measurement for metallic, pearlescent and other complex special effect color products | |||||
| Light Source | Full spectrum LED light source with blue enhancement | |||||
| Lamp Life | 5 years, 3 million times measurements | |||||
| Spectrophotometric Mode | Concave Grating | |||||
| Sensor | 256 Image Element Double Array CMOS Image Sensor | |||||
| Wavelength Range | 400nm-700nm | |||||
| Wavelength Interval | 10nm | |||||
| Measurement Range | 0~600% | |||||
| Semiband Width | 10nm | |||||
| Measuring Aperture | Illumination size Φ23mm / Measuring spot 9X12mmm (Customized: Illumination size Φ10mm / Measuring spot 6X8mmm) | |||||
| Color Space | CIE LAB,XYZ,Yxy,LCh,βxy,DIN Lab99 | |||||
| Color Difference Formula | ΔE*ab,ΔE*94,ΔE*cmc(2:1),ΔE*cmc(1:1),ΔE*00, DINΔE99,ΔE DIN6175 | |||||
| Other Colorimetric Index | Flop Index, Int-Em | Flop Index | ||||
| Observer Angle | 2°/10° | |||||
| Illuminant | D65,A,C,D50,D55,D75,F1,F2(CWF),F3,F4,F5,F6,F7(DLF),F8,F9,F10(TPL5),F11(TL84),F12(TL83/U30) | |||||
| Display | Spectrogram/Values, Samples Chromaticity Values, Color Difference Values/Graph, PASS/FAIL Result, Color Offset, Color simulation, Sample effect value, Effect difference value | Spectrogram/Values, Samples Chromaticity Values, Color Difference Values/Graph, PASS/FAIL Result, Color Offset | ||||
| Measuring Time | Approx. 1 second for one angle | |||||
| Approx. 12 seconds for all angles | Approx. 8 seconds for all angles | Approx. 6 seconds for all angles | Approx. 5 seconds for all angles | Approx. 3 seconds for all angles | Approx. 3 seconds for all angles | |
| Repeatability | Spectral reflectance: Standard deviation within 0.08% | |||||
| Chromaticity value: ΔE*ab 0.02 ( When a white calibration plate is measured 30 times at 5 second intervals after white calibration) | Chromaticity value: ΔE*ab 0.03 ( When a white calibration plate is measured 30 times at 5 second intervals after white calibration) | Chromaticity value: ΔE*ab 0.04 ( When a white calibration plate is measured 30 times at 5 second intervals after white calibration) | ||||
| Reproducibility | ΔE*<0.10,avg on the gray tile of BCRA tile set | |||||
| ΔE*<0.25,avg on the color BCRA tile set | ||||||
| Inter-instrument Error | 0.18ΔE*00(avg on reference Series II BCRA tile set) | 0.2ΔE*00(avg on reference Series II BCRA tile set) | ||||
| Effect Parameters | Sparkle Grade(SG),Diffuse coarseness(DC) and Color Variation(CV) | / | ||||
| Effect Measurement | 6 angles Sparkle Grade(SG),Color Variation(CV):15as-45°,15as-30°,15as-15°,15as15°,15as45°,15as80° | / | ||||
| 15d Diffuse coarseness(DC) | / | |||||
| Effect Repeatability | Sparkle Grade(SG) Short-term repeatability: 0.12% (10 times standard deviations) (When a color plate is measured 10 times at 10 second intervals after white calibration) | / | ||||
| Diffuse coarseness(DC) Short-term repeatability:e0.09% (10 times standard deviations) (When a color plate is measured 10 times at 10 second intervals after white calibration) | / | |||||
| Effect Reproducibility | Sparkle Grade(SG) Reproducibility: 1.9% (10 times standard deviations) (avg on reference Series II BCRA tile set) | / | ||||
| Diffuse coarseness(DC) Reproducibility: 1.4% (10 times standard deviations)(avg on reference Series II BCRA tile set) | / | |||||
| Trigger mode | Pressure sensing trigger, button trigger, software trigger | |||||
| Measuring Mode | Single measurement, average measurement (1-99), continuous measurement (1-99) | |||||
| Locating Mode | Color camera preview | |||||
| Dimension | L*W*H=195X83X128mm | |||||
| Weight | About 1Kg | |||||
| Power | lithium-ion battery, 3.7V,3200mAh, Continuous test 6000 times within 8 hours of full charge | |||||
| Display | 3.5-inch TFT color LCD, Capacitive Touch Screen | |||||
| Interface | USB, Bluetooth 5.0 | USB | ||||
| Data Storage | 1000 pcs Standards,4000 pcs Samples | |||||
| Language | Simplified Chinese, Traditional Chinese, English | |||||
| Operating Environment | 10℃ to 50℃, humidity does not exceed 85%, no condensation | |||||
| Storage Environment | -20℃ to 50℃, humidity does not exceed 85%, no condensation | |||||
| Calibration | Built-in white board parameters, external white board, black light trap, color board | Built-in white board parameters, external white board, black light trap | ||||
| Calibration Interval | 4 hours,8 hours,24 hours,Startup calibration | |||||
| Standard accessories | Power Adapter, USB Cable, User Guide,PC Software(download from the official website), Calibration Board, black light trap,Protective cap, wristband | |||||
| Optional Accessory | Micro-printer | |||||
| Notes |
Technical parameters are for reference only, subject to actual sales products. |
|||||
Multi angles spectrophotometer operation video on youtube https://www.youtube.com/watch?v=yBkKuNkgJYY